Meditech Fetal Monitor
MD900F
¡¡
|
|
Click here for
more Description
¡¡
Features |
Configuration |
Download Catalogue |
|
|
Safety
The MD900F Series complies with these
regulations and standards
IEC 60601-1,IEC 60601-1-2
FHR
measurement:
Measuring Method Ultrasonic Doppler
Measuring scope 30¡«240BPM
Measuring Error Not More Than¡À1BPM
Ultrasonic Working Frequency 2.0MHz
Ultrasonic Working Mode Pulse Diving Wave
Ultrasonic Output Intension 5mW/cm2
Display Instant FHR digital display
HR Trace 50¡«240BPM
Uterine Contractions Measurement:
Measuring Method External Measuring Method
Measuring Range 0¡«1000g
Non-linear Errorʿ5%
Display 0¡«255% unit digital display
OP trace 0¡«100 unit
FM Record Manual/automatic FM recording
hardware
configuration
Display Screen: 130mm¡Á36mm£¬320¡Á240 one-coloured dot
matrix LCD screen
Printer and printing paper: Thermal dot matrix printer
Folding thermal printing paper, 112 mm in width, 15m in
length
Size and weight Incorporate-structured device. The size
and weight for a whole machine in storage and
transportation are:
Size length X width X height 290mmX230mmX86mm
Weight 3kg
Operation environment
Operatinging power
Voltage: AC 220V¡À10%
Frequency 50Hz
Rated power 30W
Running environment
Temperature +5~+45 ¡æ
Relative humidity ¡Ü90%
Atmosphere pressure 86~106 KPa.
Saving Environment
Temperature-10~-40¡æ
Relative humidity ¡Ü80%
Without caustic gas, good ventilation
Other paremeters
FM Record Manual control button sign
Date communication interface: date transmission distance
is 2000m.
Team net mode and submachine amount: chief wire data
transmission mode
500 submachine at most
Printing paper skipping speed: 1, 2,3 cm/min, three
gears adjustable.
Printing: Real-time recording of FHR, OP traces and FM
signals
Can stop printing at set time automatically
Alarm: FHR alarm
Lower limit 80~130 adjustable
Upper limit 130~190 adjustable
Alarm triggering time(the time from FHR starting
exceeding the limit to starting alarming ) £¼2s
Alarm frequency 1 - 60 seconds adjustable.
Probe deviating alarm Triggering time < 5s
Length of probe line 1.5m - 2m
¡¡
|
|
 |
 |
|