* Display Format: LED
SPO2:35-99% in two-digit 7-segment digital;
Pulse Ratio:30-240BPM in 3-digit 7-segment
digital;
Pulse Wave: 6-segment barograph;
Battery Voltage Low Indication: LED signals.
Power Requirements:
2
AAA size 1.5V Alkaline Batteries
2 display mode. (option)
Power Consumption: less than 40mA.
Resolution: \1% for SPO2 and \1 BPM for Pulse Ratio.
Measurement Accuracy: \2% in stage of 70%-99% SPO2
and meaningless when
stage being smaller than 70%. \2 BPM for Pulse
Ratio.
Measurement Performance in Weak Filling Condition:
Oxy haemoglobin and pulse
value can be shown correctly when pulse-filling
ratio is 0.5%.
Interference Resistance Capacity against Ambient
Light: Deviation is smaller
than \1% between value of Oxy haemoglobin
measured in natural lighting indoor
condition and present lighting sources and that
measured in dark room.
It is equipped with a function switch, through which
the Oximeter can be
automatically powered off when no finger is in
the Oximeter more than 8
seconds.
|